The Microlife 47038 is a fully automatic upper arm blood pressure monitor designed for home use. It features Microlife's clinically validated AFIB detection technology, one-touch operation, a large LCD display with backlight, memory storage for multiple users, irregular heartbeat detection, WHO hypertension classification indicator, and a comfortable D-ring cuff. Below are key sections for safety, features, setup, measurement, maintenance, and troubleshooting.
Key components: Main unit with LCD display, D-ring cuff with air tube, AC adapter (optional), instruction manual, and storage case.
| Feature | Description |
|---|---|
| AFIB Detection | Clinically validated technology to screen for atrial fibrillation during measurement |
| One-Touch Operation | Automatic inflation and measurement with a single button |
| Large LCD Display | Clear backlit display showing systolic, diastolic, pulse, and date/time |
| Memory Storage | Stores up to 99 measurements per user (2 users total) |
| WHO Indicator | Color-coded bar graph showing hypertension classification |
| Irregular Heartbeat Detection | Detects and displays irregular heartbeat during measurement |
| D-Ring Cuff | Easy-to-apply cuff (22-42 cm circumference) |
| Date & Time | Automatic date and time stamp for each reading |
| Automatic Power Off | Conserves battery life by turning off after 1 minute of inactivity |
| Battery/AC Operation | 4 AA batteries or optional AC adapter |
Read all instructions before use. For accurate measurements and safe operation.
WARNING! This device is for home monitoring only, not for diagnosis or treatment. Consult your doctor for medical interpretation.
Insert batteries or connect AC adapter; set date and time.
CAUTION! Use only recommended batteries or AC adapter. Do not over-tighten cuff.
Front Panel: START/STOP button, MEMORY button, USER button (1/2), LCD display.
Display Symbols: User icon, memory index, systolic/diastolic pressure, pulse rate, date/time, battery indicator, AFIB symbol, irregular heartbeat symbol, WHO bar graph.
Settings: Press and hold MEMORY to enter setup for date/time. Press USER to switch between User 1 and User 2 memory. Press MEMORY briefly to review stored readings.
For most accurate results, measure at the same time each day.
Tip: Avoid caffeine, smoking, and exercise for 30 minutes before measurement.
The display shows systolic pressure (top number), diastolic pressure (middle number), and pulse rate.
WARNING! A single high reading does not necessarily indicate hypertension. Track trends over time.
Stores 99 readings per user (2 users total).
Note: Memory is retained when batteries are removed.
Clean with soft, dry cloth only. Do not immerse in water or use chemicals.
CAUTION! Do not disassemble or attempt repairs. No user-serviceable parts inside.
CE marked: Conforms to European Medical Device Directive 93/42/EEC. FDA Cleared: For over-the-counter home use. Manufactured by Microlife AG, Espenstrasse 139, CH-9443 Widnau, Switzerland. For support, visit www.microlife.com or contact your local Microlife service center.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace all 4 AA batteries with new ones. |
| Error E1 | Cuff not properly attached | Check tube connection; reapply cuff snugly. |
| Error E2 | Measurement error | Remain still and retry; ensure cuff at heart level. |
| Error E3 | Irregular pulse detected | Measurement may be inaccurate; rest and retry. |
| Error E4/E5 | Cuff over/under inflated | Check cuff placement; avoid movement during inflation. |
| Inconsistent readings | Incorrect posture or timing | Measure under same conditions; rest properly before. |
| AFIB symbol flashes | Possible atrial fibrillation | Consult your physician with your readings. |
Reset: Remove batteries for 1 minute to reset the device.
Service: Contact Microlife customer service or authorized dealer for repairs.