Microlife AFIB Easy Series Blood Pressure Monitor User Manual

The Microlife AFIB Easy Series is a clinically validated, automatic upper arm blood pressure monitor with integrated Atrial Fibrillation (AFIB) detection. It features a large, easy-to-read display, one-touch operation, irregular heartbeat detection, memory storage for multiple users, WHO indicator, date/time function, and risk category display. Powered by 4 AA batteries or optional AC adapter. Below are key sections for safety, features, setup, measurement, maintenance, and troubleshooting.

1. Features at a Glance 2. Safety Information 3. Package Contents 4. Device Overview 5. Initial Setup 6. Taking a Measurement 7. Understanding Results 8. Memory Function 9. Care & Maintenance 10. Technical Specifications 11. Troubleshooting

Features at a Glance

Key components: Main unit with LCD display, Easy-Wrap Cuff with D-Ring, AC adapter port, Battery compartment, START/STOP button, Memory button.

FeatureDescription
AFIB DetectionAutomatically screens for pulse irregularity suggestive of Atrial Fibrillation
One-Touch OperationAutomatic inflation and measurement with single button press
WHO IndicatorColor-coded bar graph based on WHO blood pressure classification
Multi-User MemoryStores up to 99 readings for each of 2 users
Irregular Heartbeat DetectionDetects and displays symbol for irregular heartbeats during measurement
Large DisplayEasy-to-read backlit LCD with large digits
Risk Category DisplayShows systolic/diastolic values with color indication
Date & Time StampAutomatically records date and time of each measurement
Automatic Power OffConserves battery life by turning off after 1 minute of inactivity
Easy-Wrap CuffPre-formed D-Ring cuff for easy self-application (22-42 cm)

Safety Information

Read all instructions before use. For home use only.

  1. Do not use on infants, children, or persons unable to express themselves.
  2. Consult physician before use if you have arrhythmia, pre-eclampsia, or circulatory disorders.
  3. Do not use during dialysis or if you have severe blood flow issues in measured arm.
  4. Do not inflate cuff when not wrapped around arm.
  5. Keep away from electromagnetic fields (mobile phones, microwave ovens).

WARNING! This device is not a substitute for professional medical diagnosis. AFIB detection indicates possibility of atrial fibrillation; consult physician if symbol appears frequently.

Package Contents

CAUTION! Check contents upon unboxing. Do not use if any parts are damaged or missing.

Device Overview

Front Panel: LCD Display, START/STOP button, Memory (M) button.

Display Symbols: Systolic/Diastolic pressure, Pulse rate, AFIB symbol, Irregular Heartbeat symbol, User indicator (1/2), Memory icon, Battery indicator, WHO color bar, Date/Time.

Side/Back: AC adapter socket, Cuff connector port, Battery compartment cover.

Cuff: Pre-formed tube, D-Ring, air plug, artery marker.

Initial Setup

Before first measurement, install batteries and set date/time.

  1. Insert Batteries: Open battery compartment, insert 4 AA batteries observing polarity.
  2. Set Date/Time: Press and hold M button for 3 seconds until year flashes. Use START/STOP to adjust values, M to confirm and move to next setting.
  3. Connect Cuff: Plug cuff tube into connector on side of device until it clicks.
  4. Select User: Briefly press M button to toggle between User 1 and User 2 (U1/U2 appears).

Tip: For most accurate results, wait 30 minutes after eating, smoking, or exercising before measuring.

Taking a Measurement

For reliable results, sit quietly for 5 minutes before measuring.

  1. Sit comfortably with back supported, feet flat on floor.
  2. Wrap cuff on bare upper arm. Artery marker should align with brachial artery on inner arm.
  3. Cuff edge should be 2-3 cm above elbow. Tube should run down center of arm.
  4. Rest arm on table with cuff at heart level. Relax, do not talk or move.
  5. Press START/STOP button. Device will inflate automatically.
  6. Remain still until measurement completes and cuff deflates fully.
  7. Results display with systolic, diastolic, pulse, and any detection symbols.
  8. Device automatically stores reading and powers off after 1 minute.

WARNING! Do not inflate cuff repeatedly in quick succession. Wait 2-3 minutes between measurements.

Understanding Results

Blood Pressure Values: Systolic (upper number), Diastolic (lower number) in mmHg.

WHO Indicator Colors:

Symbols:

Memory Function

Stores up to 99 readings per user with date/time stamp.

  1. View Memory: Press M button briefly to scroll through stored readings.
  2. Delete Last Reading: Press and hold M button while last reading is displayed until "CLr" appears.
  3. Delete All Readings: With device off, press and hold both START/STOP and M buttons for 3 seconds until all display segments appear, then release.

Note: Memory is maintained when batteries are replaced if done within 2 minutes.

Care & Maintenance

Clean device with soft, dry cloth only. Do not immerse in water or use chemicals.

  1. Store in provided case when not in use.
  2. Keep away from extreme temperatures, humidity, dust, and direct sunlight.
  3. Remove batteries if device will not be used for extended period.
  4. Cuff can be cleaned with mild soap and damp cloth; air dry completely before storage.
  5. Do not bend or kink cuff tube.

CAUTION! Do not attempt to repair device yourself. Contact authorized service center.

Technical Specifications

Troubleshooting

SymptomPossible CauseCorrective Action
No displayDead batteriesReplace all 4 batteries with fresh ones.
Error E1Cuff not connected properlyCheck cuff connection; ensure plug is fully inserted.
Error E2Cuff over-inflatedDeflate cuff completely and restart measurement.
Error E3Measurement errorEnsure cuff is at heart level; remain still; retry.
Error E4Pulse signal too weakReposition cuff; ensure proper fit; relax arm completely.
Inconsistent readingsMovement during measurementSit still, do not talk; wait 2-3 minutes between readings.
AFIB symbol appearsPossible atrial fibrillationConsult physician; note frequency of occurrence.
Cuff does not inflateBlocked tube or weak batteriesCheck tube for kinks; replace batteries.
Memory not workingBatteries removed too longReplace batteries within 2 minutes to preserve memory.

Reset: Remove batteries for 2 minutes, then reinsert to reset device.

Service: For technical support, contact Microlife Customer Service at 1-800-566-3456 or visit www.microlife.com