Microlife AFID ADVANCED EASY Blood Pressure Monitor User Manual
The Microlife AFID ADVANCED EASY is a fully automatic upper arm blood pressure monitor featuring AFIB (Atrial Fibrillation) detection, one-touch operation, Intellisense technology for comfortable inflation, a large LCD display with backlight, memory storage for multiple users, WHO blood pressure classification indicator, irregular heartbeat detection, date/time function, and automatic power-off. It is intended for home use to monitor systolic, diastolic blood pressure, and pulse rate. Below are key sections for safety, features, preparation, measurement, understanding results, maintenance, and troubleshooting.
Features at a Glance
Key components: Main unit with LCD display, Easy-Wrap Cuff, AC Adapter port, Battery compartment, START/STOP button, MEMORY button.
| Feature | Description |
| AFIB Detection | Advanced algorithm to screen for possible Atrial Fibrillation |
| Intellisense Technology | Automatic inflation to optimal pressure for comfort |
| One-Touch Operation | Simple START/STOP button for easy use |
| Large LCD with Backlight | Clear display of results, readable in low light |
| Dual User Memory | Stores up to 99 readings for each of two users |
| WHO Indicator | Color-coded bar graph showing blood pressure classification |
| Irregular Heartbeat Detection | Detects pulse irregularities during measurement |
| Date & Time Stamp | Records time of each measurement for tracking |
| Automatic Power Off | Conserves battery life |
| Easy-Wrap Cuff | Pre-formed cuff for simple self-application |
Safety Information
Read all instructions before use. For adult use only.
- Do not use on infants, children, or persons unable to express themselves.
- Consult your physician before use if you have arrhythmia, pre-eclampsia, or other circulatory disorders.
- This device is for measurement and monitoring only, not for diagnosis or therapy.
- Do not use in high humidity, near flammable gases, or during MRI/CT scans.
- Ensure cuff is correctly positioned; improper use may cause inaccurate readings.
WARNING! Never modify the device. Use only specified AC adapter or batteries. Keep away from children.
Parts Identification
Familiarize yourself with the device components.
- Main Unit: Houses display, controls, and internal pump.
- LCD Display: Shows systolic/diastolic pressure, pulse, icons, and memory.
- START/STOP Button: Powers on and starts measurement.
- MEMORY Button: Recalls stored measurements.
- Easy-Wrap Cuff: With D-ring and air tube connector.
- Air Tube: Connects cuff to main unit.
- AC Adapter Jack: For external power (optional).
- Battery Compartment: Holds 4 AA batteries.
Note: All parts must be intact for accurate operation.
Preparation for Measurement
For accurate results, prepare correctly.
- Rest: Sit quietly for 5 minutes before measuring.
- Position: Sit on a chair with back support, feet flat on floor.
- Cuff Placement: Apply cuff to bare upper arm. Lower edge 2-3 cm above elbow. Tube aligns with middle of arm.
- Arm Position: Rest arm on table so cuff is at heart level.
- Avoid: Talking, moving, or crossing legs during measurement.
- Timing: Measure at consistent times daily (e.g., morning and evening).
CAUTION! Do not inflate cuff when not attached to arm. Ensure cuff is not twisted.
Taking a Measurement
Follow these steps for a single measurement.
- Insert batteries or connect AC adapter.
- Wrap cuff securely on left upper arm (or arm specified by physician).
- Sit correctly as described in Preparation.
- Press the START/STOP button. All display symbols will appear briefly.
- The cuff will inflate automatically. Remain still and quiet.
- Measurement proceeds; pulse symbol flashes.
- Cuff deflates automatically. Results display with systolic, diastolic, and pulse rate.
- The device powers off automatically after 1 minute, or press START/STOP to turn off.
Tip: Wait at least 3 minutes between measurements on the same person.
Understanding the Display & Results
- Systolic Pressure: Displayed at top (e.g., 120).
- Diastolic Pressure: Displayed in middle (e.g., 80).
- Pulse Rate: Displayed at bottom (e.g., 72).
- WHO Indicator: Bar graph lights up in zone (Green=Normal, Yellow=Prehypertension, Red=Hypertension).
- User Icon (1 or 2): Shows which user's memory is active.
- AFIB Symbol: Appears if possible atrial fibrillation is detected.
- Irregular Heartbeat Symbol: Appears if pulse irregularity is detected.
- Date & Time: Displayed during memory recall.
- Low Battery Icon: Indicates batteries need replacement.
- Error Codes: (e.g., E1, E2) indicate problems (see Troubleshooting).
Memory Function
Stores readings for two users (User 1 & User 2).
- Select User: Press MEMORY button before measurement. User icon (1 or 2) blinks. Press START/STOP to confirm selection. Or, the device automatically stores to the last selected user.
- Recall Memory: Press and hold MEMORY button for 3 seconds. Display shows average of last 3 readings. Press MEMORY again to scroll through individual stored readings with date/time.
- Clear Memory: With device off, press and hold MEMORY button while inserting batteries (or switching on). All readings for the selected user are erased.
Note: Memory is retained when batteries are removed.
AFIB Detection Feature
The device analyzes pulse rhythm during measurement to screen for possible Atrial Fibrillation (AFIB).
- If the AFIB algorithm detects a pattern consistent with possible AFIB, the AFIB symbol will appear next to the measurement results.
- This is a screening feature, not a diagnosis. A positive result should be discussed with a physician.
- For accurate AFIB screening, ensure you are still and relaxed during measurement.
- The feature works in conjunction with standard blood pressure measurement; no extra steps are required.
Important: Do not rely solely on this device for diagnosing heart conditions. Always consult a healthcare professional.
Care & Maintenance
Proper care ensures longevity and accuracy.
- Cleaning: Wipe main unit and cuff with a soft, slightly damp cloth. Do not immerse in water or use solvents.
- Storage: Store in a cool, dry place. Avoid extreme temperatures, humidity, and direct sunlight.
- Cuff Care: Do not fold the cuff tightly. Keep the air tube unbent.
- Batteries: Remove batteries if device will not be used for an extended period.
- Service: Have device checked every 2 years for accuracy. Do not attempt to repair yourself.
CAUTION! Never wash or sterilize the cuff in an autoclave. Do not disassemble the unit.
Specifications
- Measurement Method: Oscillometric, fully automatic.
- Measurement Range: Pressure: 0 ~ 299 mmHg, Pulse: 40 ~ 199 beats/min.
- Accuracy: Pressure: ±3 mmHg, Pulse: ±5% of reading.
- Inflation: Automatic via Intellisense fuzzy logic.
- Cuff Size: Fits arm circumference 22 ~ 42 cm.
- Memory: 99 readings per user (2 users total).
- Power Supply: 4 x 1.5V AA batteries or DC 6V AC adapter (optional).
- Operating Conditions: 10°C ~ 40°C, 15% ~ 90% RH (non-condensing).
- Dimensions & Weight: Approx. 128mm (W) x 105mm (H) x 67mm (D), ~350g (without batteries).
Troubleshooting
| Symptom | Possible Cause | Corrective Action |
| Display is blank | Dead batteries, improper connection | Replace all 4 batteries; ensure correct polarity; check AC adapter connection. |
| Error Code E1 | Cuff not attached or leak | Ensure cuff is connected securely; check tube for damage. |
| Error Code E2 | Measurement error | Move less during measurement; ensure cuff is at heart level; retake. |
| Error Code E3 | Pulse signal too weak/irregular | Reposition cuff correctly; relax and retake measurement. |
| Inaccurate reading | Incorrect posture, recent activity | Rest for 5 minutes, sit correctly, ensure proper cuff placement. |
| AFIB symbol appears frequently | Possible arrhythmia | Consult your physician for evaluation. |
| Cuff does not inflate | Blocked air tube, weak batteries | Straighten air tube; replace batteries. |
| Memory not recalling | No data stored for selected user | Select correct user (1 or 2) before recall. |
Reset: Remove batteries for 1 minute to reset the microcomputer.
Service: For persistent issues, contact Microlife customer service or an authorized service center. Do not open the device.
Microlife AFID ADVANCED EASY Blood Pressure Monitor User Manual © - All Rights Reserved 2025