The Microlife B2 Easy is a fully automatic upper-arm blood pressure monitor designed for home use. It features one-touch operation, a large LCD display, irregular heartbeat detection, 2-user memory with 30 measurements each, WHO blood pressure classification indicator, date and time display, and automatic power-off. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit with LCD display, Upper-arm cuff with tube, 4x AA batteries, Instruction manual, Storage case.
| Feature | Description |
|---|---|
| One-Touch Operation | Fully automatic inflation and deflation |
| WHO Classification Indicator | Color-coded display of blood pressure category |
| Irregular Heartbeat Detection (IHD) | Detects pulse irregularities during measurement |
| 2x30 Memory | Stores 30 measurements for each of two users |
| Large LCD Display | Easy-to-read results with date and time |
| Automatic Power Off | Conserves battery life after 1 minute of inactivity |
| Cuff Fit Indicator | Verifies proper cuff placement |
| Average Value | Calculates average of last 3 measurements per user |
| Date & Time Stamp | Records measurement time for tracking |
| Error Detection | Alerts user to movement or incorrect positioning |
Read all instructions before use. For accurate measurements, remain still and seated.
WARNING! This device is for home monitoring only, not for diagnosis. Consult your doctor for medical interpretation.
Unpack and verify all items are present.
CAUTION! Do not use if any item is damaged. Contact retailer or Microlife service.
Front Panel: START/STOP button, Memory button (M1/M2), LCD display. Cuff: Velcro closure, air tube connector, artery marker.
Display Symbols: Systolic/Diastolic pressure, Pulse rate, User indicator (1 or 2), WHO classification bar, Irregular heartbeat symbol, Cuff fit indicator, Battery icon, Date/Time, Memory average.
Buttons: START/STOP (power on/off, start measurement), M1 (select User 1 memory), M2 (select User 2 memory).
Prepare device for first use.
Tip: Wait 30 minutes after exercise, eating, or bathing before measuring.
For accurate results, sit quietly for 5 minutes before measuring.
WARNING! Do not move or talk during measurement. Ensure cuff is not too tight or loose.
Systolic (upper value), Diastolic (lower value), Pulse rate (beats per minute).
| WHO Category | Systolic (mmHg) | Diastolic (mmHg) | Display Color |
|---|---|---|---|
| Low | < 90 | < 60 | Blue |
| Ideal | 90-119 | 60-79 | Green |
| Pre-Hypertension | 120-139 | 80-89 | Yellow |
| Hypertension | ≥ 140 | ≥ 90 | Red |
Irregular Heartbeat Symbol: Appears if irregular pulse detected during measurement. If frequent, consult physician.
Stores 30 measurements per user with date and time.
Note: Memory is retained when batteries are replaced if done within 3 minutes.
Unplug cuff tube and remove batteries before cleaning.
CAUTION! Never immerse device in water. Do not attempt to repair or disassemble.
Note: Do not use rechargeable or carbon-zinc batteries. Remove batteries if device not used for extended period.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace all 4 batteries with new alkaline batteries. |
| Error Code E1 | Cuff not connected | Ensure cuff tube is firmly inserted into device. |
| Error Code E2 | Cuff over-inflated | Remove cuff from arm, press START/STOP to release air. |
| Error Code E3 | Measurement error | Remain still, reapply cuff correctly, retry measurement. |
| Inconsistent readings | Incorrect posture/cuff position | Sit correctly with cuff at heart level. Wait 3-5 minutes between measurements. |
| Memory not working | Batteries removed too long | Replace batteries within 3 minutes to retain memory. |
Reset: Remove batteries for 3 minutes, then reinsert.
Service: For technical support, contact Microlife Customer Service or visit www.microlife.com.