The Microlife B3 Basic is a clinically validated, automatic upper arm blood pressure monitor designed for easy home use. It features one-touch operation, a large LCD display with backlight, irregular heartbeat detection, 30 memory slots, WHO blood pressure classification indicator, and a wide-range cuff (22-42 cm). Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit with LCD display, Wide-range upper arm cuff, AC adapter, Storage case, Instruction manual.
| Feature | Description |
|---|---|
| One-Touch Operation | Fully automatic inflation and deflation |
| WHO Classification Indicator | Color-coded display shows blood pressure category |
| Irregular Heartbeat Detection | Detects pulse irregularities during measurement |
| 30 Memory Slots | Stores readings with date and time |
| Large LCD with Backlight | Easy-to-read display in all lighting conditions |
| Wide-Range Cuff (22-42 cm) | Fits most adult arms |
| Automatic Power Off | Conserves battery life |
| Date & Time Stamp | Records when measurements were taken |
| Average Value Calculation | Calculates average of last 3 readings |
| Clinically Validated | Meets international accuracy standards |
Read all instructions before use. Keep for future reference.
WARNING! This device is for home use only, not for professional medical diagnosis. Consult your doctor for medical interpretation of readings.
Unpack and verify all components are present.
CAUTION! Do not use if any components are damaged. Contact customer service for replacement.
Main Unit Front: LCD display, START/STOP button. Top: Cuff connector. Rear: Battery compartment, AC adapter jack.
Display Symbols: Systolic pressure, Diastolic pressure, Pulse rate, WHO indicator, Memory icon, Irregular heartbeat symbol, Battery indicator, Date/Time, Average value icon.
Cuff: Velcro closure, air tube, artery marker (align with brachial artery).
Before first use: Install batteries or connect AC adapter, set date/time.
Tip: For most accurate readings, use AC adapter or fresh alkaline batteries.
For accurate results: Rest 5 minutes before measuring, avoid caffeine, smoking, and exercise 30 minutes prior.
WARNING! Do not move or talk during measurement. Ensure cuff is at heart level.
WHO Classification Indicator colors:
| Color | Category | Systolic/Diastolic |
|---|---|---|
| Green | Optimal | <120 / <80 mmHg |
| Yellow | Normal | 120-129 / 80-84 mmHg |
| Orange | High Normal | 130-139 / 85-89 mmHg |
| Red | Hypertension | ≥140 / ≥90 mmHg |
Irregular Heartbeat Symbol: Appears if irregular pulse detected during measurement. If symbol appears frequently, consult your physician.
Error Codes: E1 (cuff not properly applied), E2 (measurement error), E3 (pulse rate out of range), E4 (excessive movement).
Device stores last 30 readings with date and time.
Note: Memory is maintained for approximately 72 hours when batteries are removed.
Unplug AC adapter/remove batteries before cleaning.
CAUTION! Do not immerse in water. Do not autoclave. Do not use near flammable gases.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace all 4 batteries or use AC adapter. |
| Error E1 appears | Cuff not connected properly | Reconnect cuff securely; ensure no kinks in tube. |
| Inconsistent readings | Movement during measurement | Remain still and quiet; retake measurement. |
| Cuff doesn't inflate | Air leak | Check cuff/tube connections; replace cuff if damaged. |
| Memory not working | Batteries removed too long | Memory clears after >72 hours without power. |
| Display dim | Low battery | Replace batteries or use AC adapter. |
Reset: Remove batteries for 30 seconds, then reinstall.
Service: Do not attempt to repair. Contact Microlife customer service or authorized service center.
Support: Microlife AG, Espenstrasse 139, CH-9443 Widnau, Switzerland. Website: www.microlife.com. US Support: 1-800-566-0456.