The Microlife MSC Blood Pressure Monitor is a clinically validated, automatic upper-arm blood pressure monitoring device featuring Microlife Average Mode (MAM) technology for accurate measurement, large LCD display with backlight, irregular heartbeat detection, WHO hypertension indicator, memory storage for multiple users, adjustable cuff, date/time display, and battery/power adapter operation. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit with LCD display, Adjustable upper-arm cuff, AC power adapter, User manual, Storage case.
| Feature | Description |
|---|---|
| MAM Technology | Microlife Average Mode for clinical accuracy |
| WHO Hypertension Indicator | Color-coded display based on WHO categories |
| Irregular Heartbeat Detection | Detects pulse irregularities during measurement |
| Multi-User Memory | Stores readings for up to 2 users (99 readings each) |
| Large LCD with Backlight | Easy-to-read display in all lighting conditions |
| Adjustable Cuff | Fits arm circumferences 22-42 cm |
| Date & Time Stamp | Records measurement date and time |
| Automatic Power Off | Conserves battery life |
| Battery/AC Operation | 4xAA batteries or included power adapter |
| Average Calculation | Calculates average of last 3 readings |
Read all instructions before use. Consult physician for interpretation of readings.
WARNING! This device is for home use only, not for clinical diagnosis. Seek medical advice for abnormal readings.
CAUTION! Check all items upon unpacking. Contact retailer if anything is missing or damaged.
Front Panel: START/STOP button, MEMORY button, USER SELECT button, LCD display.
Display Symbols: Systolic pressure, Diastolic pressure, Pulse rate, WHO indicator, User icon, Memory icon, Battery indicator, Irregular heartbeat symbol, Date/Time, Average icon.
Connections: Cuff tube connector, DC power jack.
Prepare device for first use.
Tip: For most accurate readings, use AC power adapter when available.
For accurate results, follow these steps carefully.
WARNING! Do not inflate cuff repeatedly in quick succession. Wait 2-3 minutes between measurements.
The display shows systolic pressure (top number), diastolic pressure (middle number), and pulse rate (bottom number).
| WHO Category | Systolic (mmHg) | Diastolic (mmHg) | Display Color |
|---|---|---|---|
| Optimal | < 120 | < 80 | Green |
| Normal | 120-129 | 80-84 | Green |
| High Normal | 130-139 | 85-89 | Yellow |
| Hypertension Grade 1 | 140-159 | 90-99 | Orange |
| Hypertension Grade 2 | 160-179 | 100-109 | Red |
| Hypertension Grade 3 | ≥ 180 | ≥ 110 | Red (flashing) |
Irregular Heartbeat Symbol: Appears if irregular pulse is detected during measurement. Consult physician if this appears frequently.
Device stores 99 readings for each of 2 users.
Note: Memory is preserved when batteries are changed if done within 3 minutes.
Unplug power adapter and remove batteries before cleaning.
CAUTION! Do not immerse any part of the device in water. Do not autoclave.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace all 4 batteries; check polarity |
| Error code E1 | Cuff not properly attached | Check cuff connection; reattach securely |
| Error code E2 | Measurement error | Reposition cuff at heart level; remain still |
| Error code E3 | Pulse too irregular | Wait 5 minutes; retry measurement |
| Inaccurate reading | Wrong cuff position | Ensure cuff is 1-2 cm above elbow on bare skin |
| Cuff doesn't inflate | Tube blockage/kink | Check tube for kinks; ensure proper connection |
| Memory lost | Batteries removed too long | Replace batteries within 3 minutes of removal |
| Irregular heartbeat symbol | Pulse irregularity detected | Consult physician if symbol appears frequently |
Reset: Remove batteries and power adapter for 5 minutes, then reinsert.
Service: Do not attempt to repair. Contact Microlife customer service or authorized service center.